* Insert the " 1.0 ppm standard solution into the

" Container " ( 3-2 , Fig. 1) and envelope in the
" Cover " ( 3-1, Fig. 1 ) completely, other
procedures please refer to chapter 4-1, 4-2.

Press " CAL Button " ( 3-5, Fig. 1 ) once, the
Display will show following text with flashing ( approx.
10 seconds ).

flashing

CAL

1.00

ppm

Then the LCD display will return to normal
measurement screen, now the meter is finished the
calibration procedures ( Zero, 1.0 ppm calibration )
completely and ready for the measurement.

Remark :

The calibration procedures for the function of " Total

chlorine " and the " Free chlorine " are independent.

The complete calibration procedures for the function

of " Free chlorine " measurement should be

executed by

* Zero Chlorine standard solution

* 1.0 ppm Free standard solution

The complete calibration procedures for the function
of " Total chlorine " measurement should be
executed by

* Zero Chlorine standard solution
* 1.0 ppm Total standard solution