beautypg.com

Specifications – Sper Scientific LUCL-2006 Chlorine Meter User Manual

Page 4

background image

2. SPECIFICATIONS

Circuit

Custom one-chip of microprocessor LSI
circuit.

Display

LCD size : 41 mm x 34 mm

Range

Free chlorine ( CL ) :
0.00 to 3.50 ppm ( mg/L ).
Total chlorine ( CL ) :
0.00 to 3.50 ppm ( mg/L ).

Resolution

0.01 ppm ( mg/L ).

Accuracy

± 0.02 ppm ( mg/L ).

@ 1.00 ppm ( mg/L )

Light source

LED, 525 nm.

Light detector Photo diode
Method

The measuring method is an adapta-
tion of the USEPA Method 330.5 and
Standard Method 4500-Cl G.

* The reaction between free ( total )

chlorine and the DPD reagent cause
a pink tint in the sample.

Response time Less than 10 seconds.
Sample volume 10 mL.
Data Hold

Freeze the display reading.

Memory Recall Maximum & Minimum value.
Display

Approx. 1 second.

Sampling
Time
Power off

Auto shut off saves battery life or
manual off by push button.

Calibration Zero

chlorine.

points

1.00 ppm ( Free chlorine ).
1.00 ppm ( Total chlorine ).

2