Unit

HygroPalm HP22-A hand-held indicator:

User Guide

Instruction Manual

Document Type

Page 15 of 16

Document title


© 2010-2012; Rotronic AG E-M-HP22-V2_13

Service connector

HP22

Interface type

UART

(Universal Asynchronous Receiver Transmitter)

Maximum service cable length

5 m (16.4 ft)

General specifications

HP22

Display

LC, 1 or 2 decimals resolution, backlight, trend, alarm and low battery indication

Probe material

Polycarbonate

Probe dust filter material

Polyethylene

Housing material

ABS

Housing protection grade

IP 40

Overall dimensions

270 x 70 x 30 mm (10.63 x 2.76 x 1.17”)

Probe dimensions

80 x 15 mm (3.15 x 0.59”)

Weight

About 198 g (7.0 oz)

Conformity with standards

HP22

CE / EMC immunity

EMC Directive 2004/108/EG: EN 61000-6-1: 2001, EN 61000-6-2: 2005
EN 61000-6-3: 2005, EN 61000-6-4: 2001 + A11

Solder type

Lead free (RoHS directive)

FDA / GAMP directives

compatible

Environmental limits

HP22

Storage and transit

-20…+70 °C / 0…100 %RH, non condensing

Operating limits at electronics

-10….60 °C (limited by LC display)
0…100 %RH, non condensing

Temperature limits at probe

Depends on probe model

Maximum humidity at probe

100 %RH up to 80 °C (176 °F)
75 %RH at 100 °C (212 °F)
45 %RH at 125 °C (260 °F)
15 %RH at 150 °C (302 °F)

Maximum air velocity at probe

20 m/s (3,935 ft /min)

Critical environments

Humidity sensor: as per DV04-14.0803.02 - Critical chemicals

7.2 Dew point accuracy

See document E-M-HC2 Probes > Dew point accuracy

8

Accessories

For accessories and parts such as the HW4 configuration software, service cables, calibration accessories
and spare dust filters, please see document E-M-HC2-accessories