2 dew point accuracy, 8 accessories, Dew point accuracy – ROTRONIC HP22A User Manual
Page 15: Accessories, 8accessories, Hygropalm hp22-a hand-held indicator: user guide
E-M-HP22-V2_13
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroPalm HP22-A hand-held indicator:
User Guide
Instruction Manual
Document Type
Page 15 of 16
Document title
© 2010-2012; Rotronic AG E-M-HP22-V2_13
Service connector
HP22
Interface type
UART
(Universal Asynchronous Receiver Transmitter)
Maximum service cable length
5 m (16.4 ft)
General specifications
HP22
Display
LC, 1 or 2 decimals resolution, backlight, trend, alarm and low battery indication
Probe material
Polycarbonate
Probe dust filter material
Polyethylene
Housing material
ABS
Housing protection grade
IP 40
Overall dimensions
270 x 70 x 30 mm (10.63 x 2.76 x 1.17”)
Probe dimensions
80 x 15 mm (3.15 x 0.59”)
Weight
About 198 g (7.0 oz)
Conformity with standards
HP22
CE / EMC immunity
EMC Directive 2004/108/EG: EN 61000-6-1: 2001, EN 61000-6-2: 2005
EN 61000-6-3: 2005, EN 61000-6-4: 2001 + A11
Solder type
Lead free (RoHS directive)
FDA / GAMP directives
compatible
Environmental limits
HP22
Storage and transit
-20…+70 °C / 0…100 %RH, non condensing
Operating limits at electronics
-10….60 °C (limited by LC display)
0…100 %RH, non condensing
Temperature limits at probe
Depends on probe model
Maximum humidity at probe
100 %RH up to 80 °C (176 °F)
75 %RH at 100 °C (212 °F)
45 %RH at 125 °C (260 °F)
15 %RH at 150 °C (302 °F)
Maximum air velocity at probe
20 m/s (3,935 ft /min)
Critical environments
Humidity sensor: as per DV04-14.0803.02 - Critical chemicals
7.2 Dew point accuracy
See document E-M-HC2 Probes > Dew point accuracy
8
Accessories
For accessories and parts such as the HW4 configuration software, service cables, calibration accessories
and spare dust filters, please see document E-M-HC2-accessories