Default output range, Formulas for calculation of output values, Output configurations – ROTRONIC CF8-W-Disp-AL User Manual
Page 3
Dokument
UMA 145
Rev
3
Page
3 (8)
Terminal
Default
Output
Default Output
Range
Outputs of this
sensor
Output Range of this
sensor
OUT1
0-20 mA
0-4 % CO
2
OUT2
4-20 mA
0-4 % CO
2
Relay 1
Closed
Open
> 1,5 % CO
2
< 1,4 % CO
2
Relay 2
Closed
Open
> 3,0 % CO
2
< 2,9 % CO
2
Table I. Default output configurations for CF8-W-Disp-AL
Formulas for calculation of output values
Configuration
Output Range
Formula for calculation
0-10 VDC
0-4 % CO
2
CO
2
value = Volt/10*4
2-10 VDC
0-4 % CO
2
CO
2
value = (Volt –2)/8*4
4-20 mA
0-4 % CO
2
CO
2
value = (mA-4)*4/16
a-b
c–d
Value = (reading-a)/(b-a)*(d-c)+c
a = lowest value of the configuration
b = highest value of the configuration
c = lowest value of the range
d = highest value of the range
Table II. Calculation of CO
2
value for CF8-W-Disp-AL
Output Configurations
If other options are needed for the application, the output jumpers have to be
configured before the unit is powered up. Each jumper selection is independent from
the others, except for the “Start point selection” jumper, which affects both, Out(1) &
Out(2), linear outputs. Alternative measuring ranges of the outputs can be selected
with the software UIP (version 4.3 or later).