Estimated non-linearity of kd5100 – Kaman KD-5100 User Manual
Page 32

32
Non-Linearity
Non-linearity is computed as the maximum error from a best fit (least squares) line and divided
by the full range. A system with a non-linearity of
±
0.2%FR and a
±
10 mil (
±
0.25mm) range (20
mil (0.5mm) Full Range) would have a maximum deviation of 0.2%x20mil=0.04mils
(0.2%x0.5mm=0.001mm).
Estimated Non-Linearity of KD5100
0.00%
1.00%
2.00%
3.00%
4.00%
5.00%
6.00%
0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100
%
Full Range as % of Coil Diameter
B
e
st
Fi
t Non-
Li
n
e
a
ri
ty
a
s
%
of
F
ul
l Ra
nge
Below is the same graph on an expanded scale:
Estimated Non-Linearity of KD5100
0.00%
0.10%
0.20%
0.30%
0.40%
0.50%
0.60%
0.70%
0.80%
0.90%
0%
10%
20%
30%
40%
50%
Full Range as % of Coil Diameter
Bes
t F
it
N
on-
L
ine
ar
it
y
as
%
of
F
u
ll
Range