beautypg.com

Dwyer LDL2-030 User Manual

Dwyer instruments, inc, Model ldl2-030 vented water level data logger

background image

Bulletin DA-LDL2-030

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

Specifications - Installation and Operating Instructions

Model LDL2-030 Vented Water Level Data Logger

The Model LDL2-030 Vented Water Level Data Logger has a vented

cable and end cap to allow it to automatically compensate for atmospheric

pressure fluctuations. The data logger measures water level up to 30 ft

(9.14 m) and has an internal temperature sensor that accurately records

temperatures between -40 to 176°F (-40 to 80°C). The innovative design

makes it ideal for use in waste water treatment facilities as well as

monitoring well and ground water levels. The battery-powered data logger

can store over 16,000 measurements per channel, and the software allows

for user selectable sampling rates. The easy to use DL700 software allows

the LDL2-030 to be easily started and stopped from a PC or delayed to

start up to six months in advance.

Before the logger is submerged, make sure all caps and parts of the data

logger are tightened to ensure that it is properly sealed. Since the LDL2-

030 is atmospheric pressure compensated, the communications end of

the data logger, (the end with the vented cap and the key ring hole) must

remain above water. Do NOT allow water or other liquids to come into

contact with this end. Tie a secure line to the key ring hole before

submerging.

SOFTWARE INSTALLATION

1. Insert installation CD and click Install Dwyer 2.00.

2. Follow onscreen instructions.

3. Next, click Install USB Interface Drivers on the CD Contents screen.

4. Click Install.

SPECIFICATIONS

Range: Temperature: -40 to 176°F (-40 to 80°C); Water Level: 0 to 30 ft

(0 to 9.14 m).

Memory Size: 16,383 readings per channel.

Accuracy: ±0.9°F (±0.5°C), ±0.3%FSR @ 77°F (25°C).

Resolution: Temperature: 0.2°F (0.1°C); Water Level: 0.02 in (0.051 cm).

Sampling Method: Stop on memory full.

Sampling Rate: Selectable from 2 sec to 12 hrs.

Computer Requirements: Windows

®

95, Windows

®

98, Windows

®

2000, Windows

®

ME, Windows NT

®

, and Windows

®

XP operating

systems, one free USB port.

Power Requirements: User replaceable, 3.6V lithium battery.

Battery Life: 1 year (approx).

Interface: USB port (interface cable required).

Material: 303 stainless steel.

Weight: 3 lb (1.4 kg).

Agency Approvals: CE.

A message may appear stating that the software has not passed

Windows

®

Logo Testing. This software has been tested and works

properly. Click Continue Anyway if this window appears.

Windows

®

, Windows NT

®

, and Windows Vista

®

are registered trademarks of

Microsoft Corporation.

NOTICE

NOTICE

8-53/61

[224.41]

7-1/32

[178.44]

1-1/4

[31.75]

SENSOR END

CONNECTED TO

RECEIVER END BY

30 FOOT CABLE

1 [25.40]

CONNECTED TO

SENSOR END BY

30 FOOT CABLE

RECEIVER END

6-53/64

[173.33] 4-19/32

[116.84]

DA-LDL2-030:TEMPLATE 11/24/08 10:30 AM Page 1