LCD INDICATORS
BUTTON OPERATION
The button performs two functions:
1. To start the logger when in push to start mode.
2. To enable the display and cycle through the available display modes (see diagram
below).
If the button is not pressed for 30 seconds. The LCD disables and the logger returns
to the beginning of the display cycle.
BATTERY INSTALLATION
Before using the data logger, you will need to insert the 3.6 V 1/2 AA battery provided,
following the instructions below.
1. Use screwdriver to gently depress tab of the battery compartment.
2. Pull battery cover towards you, away from the unit.
3. Install the battery, taking note of polarities.
4. Replace battery cover. Cover is secure when tab returns to original position.
SOFTWARE INSTALLATION
This software must be installed prior to the USB Driver. Please follow the instructions
below then refer to the USB Driver Installation based on the Windows
®
operating
system you are using.
1. Visit our website www.dwyer-inst.com/DW-USBSoftware and click on the
download link. If the download folder doesn’t open directly, go to your computer’s
“Downloads” folder and double-click on the “EasyLogUSBV610_Dwyer” folder.
2. Double-click on “setup” then follow the on screen instructions to complete the
Configuration Software Installation.
3. An EasyLog USB icon will now be placed on your desktop.
USB DRIVER INSTALLATION
Windows
®
XP/Windows Vista
®
:
You must install the Configuration Software as instructed in the previous section
before going through these steps.
1. Insert the USB data logger directly into a USB port on the computer. Installation
screens should automatically appear.
2. Follow the on screen instructions to complete the Configuration Software
Installation. If a compatibility error window appears, click “Continue Anyway”. This
will not cause any adverse changes/effects on your computer.
3. You can now begin using our data logger by double clicking on the EasyLog USB
icon on your computer’s desktop.
Windows
®
2000:
You must install the Configuration Software as instructed in the previous section
before going through these steps.
1. Insert the USB data logger directly into a USB port on the computer. Installation
screens should automatically appear.
2. Follow the on screen instructions to complete the USB Driver Installation.
3. You can now begin using our data logger by double clicking on the EasyLog USB
icon on your computer’s desktop.
MAINTENANCE
Upon final installation of the Compact DW-USB-LCD Data Logger, no routine
maintenance is required. Leaving the logger plugged into your computer for an
extended period of time will considerably lower its battery life. A periodic check of
system calibration is recommended. The Series DW-USB-LCD is not field serviceable
and should be returned if repair is needed (field repair should not be attempted and
may void warranty). Be sure to include a brief description of the problem plus any
relevant application notes. Contact customer service to receive a return goods
authorization number before shipping.
Last
temperature reading
Highest logged
temperature reading
Highest logged
humidity reading
Lowest logged
humidity reading
Lowest logged
temperature reading
Last logged
humidity reading
1
2
3
4
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000
www.dwyer-inst.com
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Fax: 219/872-9057
e-mail: [email protected]
©Copyright 2013 Dwyer Instruments, Inc.
Printed in U.S.A. 9/13
FR# R6-443584-50 Rev. 1
Temperature
scale
Relative Humidity
symbol
Minimum logged
value symbol
Maximum logged
value symbol
Reading
Negative symbol
Temperature scale symbols
Maximum/Minimum
logged value symbol
LCD INDICATION
Display
Logger Status
Explanation
Delayed Start
Push to Start
Logging
Stopped
This is shown when the logger is set to
start at a specific date and time*
This is shown when the logger is setup for
“Push to start” logging
This is shown when the logger is running
in “LCD off” mode, and the button is
pressed. The display clears again after
three seconds.
If the logger has not been set to log and
the button is pressed, three dashes are
displayed for three seconds
Otherwise the display will remain blank.