Dwyer TPT User Manual
Dwyer instruments, inc, Series tpt pressure transmitter
Series TPT Pressure Transmitter
Specifications - Installation and Operating Instructions
Bulletin P-TPT
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000
www.dwyer-inst.com
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Fax: 219/872-9057
e-mail: [email protected]
SPECIFICATIONS
Service: Compatible liquids and gases.
Wetted Materials: 17-4PH SS.
Accuracy: ±1% FS (RMS).
Stability: < ±0.25% FS per year.
Temperature Limits: -13° to 185°F (-20° to 85°C).
Compensated Temperature Range: -13° to 185°F (-20 to 85°C).
Pressure Limits: Proof pressure: 2x FS; Burst pressure: 3x FS.
Thermal Effect:
Zero shift: ±0.01% FS per 1°C;
Span error: ±0.02% FS per 1°C.
Power Requirements: See table.
Output Signal: See table.
Loop Resistance: < 100Ω.
Electrical Connection: Packard connector.
Process Connections: 1/4˝-18 male NPT.
Shock: ±20 g.
Vibration: 50 g @ 20 to 2000 Hz.
Weight: 0.20 lb (0.09 kg).
The Series TPT Pressure Transmitter is the ideal choice for all types of industrial
pressure applications. Machined from a solid piece of 17-4PH SS, the TPT
provides a leak proof, all metal system without O-rings, welds, or organics exposed
to the pressure media. Its design allows for stable operation when subject to shock
and vibration. Utilizing piezoresistive technology with digital compensation and
temperature correction, the TPT gives high accuracy and stability under harsh
environmental conditions. Available in ratiometric and 4 to 20 mA outputs.
FEATURES
• Robust construction to withstand high vibration
• 17-4PH SS wetted material
• Standard packard connector
APPLICATIONS
• Process automation and control
• Test and measurement equipment
• Factory automation
• Energy management
3-3/8
[85.88]
31/64
[12.15]
35/64
[13.72]
7/8
[22.23]
Do not exceed specified supply voltage ratings. Permanent
damage not covered by warranty will result. This device is not
designed for 120 or 240 volt AC operation. Use only 4.75 to 5.25 volts DC for
ratiometric outputs and 9 to 30 volts DC for 4 to 20 mA outputs.
CAUTION
Pressure Range
0 to 100 psig
0 to 250 psig
0 to 500 psig
0 to 1000 psig
0 to 2500 psig
0 to 5000 psig
Maximum Pressure
200 psig
500 psig
1000 psig
2000 psig
5000 psig
10000 psig
Burst Pressure
300 psig
750 psig
1500 psig
3000 psig
7500 psig
15000 psig
Pressure Ranges
Model
TPT-R01
TPT-R02
TPT-R03
TPT-R04
TPT-R05
TPT-R06
TPT-C01
TPT-C02
TPT-C03
TPT-C04
TPT-C05
Range psi (bar)
100 (6.89)
250 (17.24)
500 (34.47)
1000 (68.95)
2500 (172.37)
5000 (344.74)
100 (6.89)
250 (17.24)
500 (34.47)
1000 (68.95)
2500 (172.37)
Power Requirements (VDC)
4.75 to 5.25
4.75 to 5.25
4.75 to 5.25
4.75 to 5.25
4.75 to 5.25
4.75 to 5.25
9 to 30
9 to 30
9 to 30
9 to 30
9 to 30
Output Signal
0.5 to 4.5 VDC
0.5 to 4.5 VDC
0.5 to 4.5 VDC
0.5 to 4.5 VDC
0.5 to 4.5 VDC
0.5 to 4.5 VDC
4 to 20 mA
4 to 20 mA
4 to 20 mA
4 to 20 mA
4 to 20 mA