Dwyer DM-1100 User Manual

Page 4

background image

MAIN MENU PROGRAMMING INSTRUCTIONS
(Upper Right Display Reads MENU)

Security Menu: SECr-MENU
Lock out access to all menus or lock out access to all menus except Auto-Zero
function (Auto-ZERO).

Operation Menu: OPEr-MENU
Select the measurement type – pressure, velocity or flow and corresponding
engineering units (Velocity and flow are not available for Model DM-1100).

Display Menu: diS-MENU
Monitor and adjust display related settings: Filter type, filter point, dampening,
display update, peak and valley.

Advanced Menu: AdU-MENU
Perform auto-zero or full-scale calibration.

SECURITY MENU: SECR-MENU
With the gage reading pressure (home position), press and hold the MENU key
until the Security MENU is displayed. Press the E key to show the security SUB
MENU SECr–0001. When the security SUB MENU is selected, the present
security level is displayed in the upper right hand display. To change the security
level, adjust the number displayed to the number shown in the following table for
the desired security level.

The password values shown in the table cannot be altered, so retain a copy of
these pages for future reference.

OPERATION MENU: OPER-MENU
With the gage reading a numerical value (home position), press and hold the
MENU key until the Security MENU is displayed. Press the q key to show the
OPEr-MENU. Press the E key to show the Operation SUB MENUS.

The Operation SUB MENUS for Series DM-1100 is PrES–INWC (Pressure–Unit)

PRESSURE SUB MENU: PRES – INWC
PrES – INWC: 
Pressure – inches of water column

With the LCD reading PrES – INWC, press the E key; the display will blink. Press
the q key to change the pressure unit then press the E key to save the desired
unit.

For pressure measurement, the following units are available:
INWC – Inches of water column

MBAR – Millibar

PSI – Pounds per square inch

INHG – Inches of mercury

KPA – Kilopascals

MMHG – Millimeters of mercury

PA – Pascals

FS – % of full scale

MMWC – Millimeters of water column

Security Level
Displayed
1
2

3

Access

All menus access
All menus locked except
the auto-zero function
All menus locked

Password Value
to Enter
10
90

111

TABLE 1 PRESSURE RANGE VS. AVAILABLE UNITS

NOTE: OFL (over flow) or UFL (under flow) will appear when the low and high
range levels have been exceeded by 10% F.S.

in w.c.
0.250
0.500
1.000
2.000
5.000
10.00
15.00
25.00
50.00
100.0

psi




0.181
0.361
0.543
0.903
1.806
3.613

kPa
0.062
0.124
0.249
0.498
1.245
2.491
3.738
6.227
12.45
24.91

Pa
62.20
124.5
249.1
498.2
1245
2491
3738
6227

mbar
0.622
1.245
2.492
4.982
12.45
24.91
37.38
62.27
124.5
249.1

mm w.c.
6.35
12.70
25.40
50.80
127.0
254.0
381.0
635.0
1270
2540

in Hg




0.368
0.736
1.104
1.839
3.678
7.355

mm Hg
0.467
0.934
1.868
3.736
9.34
18.68
28.02
46.71
93.42
186.8

% of
FS
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0

Resolution
(in w.c.)
0.001
0.001
0.001
0.001
0.002
0.010
0.010
0.010
0.020
0.100