Overview, Introduction, Capabilities – Crystal ConfigXP Digital Pressure Gauge Software User Manual
Page 3
![background image](/manuals/371536/3/background.png)
Overview
3
Confi gXP User’s Manual
Overview
—
INTRODUCTION
Confi gXP provides an easy way to customize your gauge (DTG, XP2, or XP2i). With only a serial cable (P/N: 2400, or equivalent*), a computer running Windows,
and the Confi gXP program you can:
•
View and print an
as receive versus as left change report
•
Load and save custom confi gurations
•
Return the gauge to factory confi guration
•
Select pressure units (e.g., disable inches Hg, if you never use this scale)
•
Adjust (calibrate) the gauge
•
Identify fi rmware version
Note:
DTGs require an RS-232 interface adapter kit (P/N: 2746).
—
CAPABILITIES
Some features are dependant on the model and fi rmware version of your device. You may update your fi rmware
.
For many gauges, you will be able to:
•
Password protect the gauge to prevent unauthorized changes
•
Disable (or enable) the
(
peak
)
button
•
Select your preferred H
2
O water density: 4°C, 60°F, 68°F/20°C
•
Defi ne a special pressure unit (e.g., display pressure as torque)
•
Store a message or identifi cation number
•
Limit or expand the zero adjustment range
•
Program a TARE value
•
Enable and adjust display damping (averaging)
•
Display rate of pressure change
•
Disable the ability to adjust (calibrate) the gauge from the keypad
Note:
Features not available on your gauge will be grayed out in Confi gXP.
!
CAUTION: Straight pass-through RS-232 cable required. DO NOT use a null-modem cable.
!
CAUTION: Changing the settings of your gauge using this program can cause the unit to display inaccurate readings or behave diff erently than the
factory defaults. Only personnel qualifi ed to make changes should do so.