Password protection, Loading and saving configurations, Reset to factory defaults – Crystal ConfigXP Digital Pressure Gauge Software User Manual
Page 12: Loading and saving confi gurations
Settings
12
Confi gXP User’s Manual
—
PASSWORD PROTECTION
The password protection feature of the gauge allows you to prevent unauthorized changes to the gauge confi guration by setting a four digit password. Once
the password is set you will not be able to make any changes to the gauge through Confi gXP without the password. In addition, any functions within the
gauge itself that can be accessed by the keypad or serial interface, such as water reference select or userspan adjustment, will be disabled. Any serial interface
command that attempts to change a protected feature will return a
X,0
message indicating that feature cannot be changed. Passwords can only be set or
removed using Confi gXP.
In the event you lose the password, you will need to contact the factory for an unlock code, which will remove the password protection.
—
LOADING AND SAVING CONFIGURATIONS
Use the
Load Confi guration
and
Save Confi guration
buttons to create custom setups–like enabling a limited set of engineering units or enabling other
features in the gauge–and storing those confi gurations for later use. Applications might include using a saved confi guration to setup multiple gauges identi-
cally, or using diff erent confi gurations for the same gauge to customize it for diff erent applications. Confi gurations do not include the password or userspan
value.
—
RESET TO FACTORY DEFAULTS
The
Load Factory Defaults
button will return the gauge to the confi guration as shipped from the factory. As part of the reset function, the password will be
removed from the gauge, but the userspan value will be unaff ected.