Functions

4

XP2i Operation Manual

clear

zero

resetting (clearing) recorded Peak Values

Peak values can only be cleared when displaying either a Peak High or Peak Low recorded pressure. Press the

(

clear

)

button for at least ½ second. Dashed

lines will briefly appear across the display indicating that both Peak values have been cleared. Both Peak High and Peak Low values will then display the cur-
rent applied pressure. Pressing the

(

clear

)

button while either the Peak High (

HI

) or Peak Low (

LO

) icon is displayed will not affect the zero value. If you

need to rezero the gauge, you must turn off both peak icons by pressing the

(

peak

)

button.

clear

zero

Zero

If you attempt to zero the gauge while applying a pressure which exceeds the Zero Limit (set in ConfigXP, defaults to 20 psi),

the command will be ignored and “--HI-” will be displayed.

X

to Zero the XP2i

Turn off peak indication by pressing the

(

peak

)

button repeatedly until the HI and LO icons are off, then press the

(

zero

)

button for at least ½ second when

the gauge is vented to atmosphere.

The display will then briefly flash all dashed lines (

), indicating that it has been re-zeroed. Absolute gauges will now indicate gauge pressure.

!

WaRninG: this gauge can display zero pressure when connected to a source of pressure! do not rely on the display indication before disconnecting—

it may not be indicating true pressure. never disconnect pressure instrumentation without first relieving system pressure!

X

to clear the Zero Value on an XP2i

Turn off peak indication as described above, then press and hold the

(

zero

)

button until the display changes from (

) to (

).

This is especially useful for absolute gauges that have been zeroed to use for gauge pressure measurement.