Performance, Reference conditions, Performance under reference conditions – West Control Solutions N8080 (for 1xxx) User Manual
Page 39: Operating conditions

A.9
PERFORMANCE
Reference Conditions
Ambient Temperature:
20
°
C
±
2
°
C
Relative Humidity:
60 - 70%
Supply Voltage:
90 - 264V AC 50Hz
Source Resistance:
<10
Ω
for thermocouple input
Lead Resistance:
<0.1
Ω
/lead balanced (Pt100)
Performance Under Reference Conditions
Common Mode Rejection:
>120dB at 50/60Hz giving negligible effect
at up to 264V 50/60Hz.
Series Mode Rejection:
>500% of span (at 50/60Hz) causes
negligible effect.
Thermocouple Inputs
Measurement Accuracy:
±
0.05% typical,
±
0.1% maximum, of span
±
1LSD. NOTE: Reduced performance with
Type “B” Thermocouple below 600
°
C
(1112
°
F).
Linearisation Accuracy:
High resolution ranges - better than
±
0.5
°
C
Low resolution ranges - better than
±
0.05
°
C
Cold Junction
Compensation:
±
0.3% typical,
±
0.5% maximum.
RTD Inputs
Measurement Accuracy:
±
0.02% typical,
±
0.06% maximum, of span
±
1LSD
Linearisation Accuracy:
High resolution ranges - better than
±
0.5
°
C.
Low resolution ranges - better than
±
0.05
°
C.
Operating Conditions
Ambient Temperature:
0
°
C to 55
°
C
PM090-A
Volume II
A-5
59135