3.5. Source – Voltage

The unit outputs voltage from 0 to 22V in three ranges, 0 to 0.22V (displayed in mV); 0 to
2.2V and, 0 to 22V.

Set function switch to SRC

Set OFF/V/mA switch to V

Set range switch to the desired range

Set output switch to NORM

Set multi-turn output control to minimum (anticlockwise)


Connect the unit to the process instrument

observing correct polarity and adjust the
multi-turn output control for coarse control and
INC and DEC buttons for fine control.

NOTE:

1) The maximum output current is limited to approx 50mA.

2) The units output resistance is approx 2.5 ohms. Therefore the user should be aware of the loading
implications - a load of 2.5K ohms will produce a 0.1% error in the output voltage. To maintain full
accuracy the maximum load should be restricted to 1.1K ohms.

3.6. Source – Current

The unit outputs current from 0-22mA in three selectable ranges, 0 to 0.22mA (displayed in
uA), 0 to 2.2mA and 0 to 22mA.

• Set function switch to SRC
• Set OFF/V/mA switch to mA
• Set range switch to the desired range
• Set output switch to NORM
• Set multi-turn output control to minimum (anticlockwise)

Connect the unit to the process instrument
observing correct polarity and adjust the
multi-turn output control for coarse control and
INC and DEC buttons for fine control.

-

+