1048 Technical Manual

7.

Maintenance

7.1. Recalibration

Calibration equipment required:

A) A precision DC voltage calibrator with an accuracy of 0.01% or better. Examples of
suitable instruments are Time Electronics’ 1017, 5018, 5025 or 5051.

B) DMM with accuracy of 0.01% or better. Examples of suitable instruments are Time
Electronics’ 5075, or 5065.

Note: It is advisable to use leads with low thermal emf connections. This will ensure that
stray thermal emfs do not cause errors when calibrating the low voltage range (.22V) which
has a setting resolution of 10uV.






The unit connected to a Time
Electronics’ 1017 Precision
0.005% DC V & I calibrator.

The unit connected to a Time
Electronics 5065 6.5 digit
Precision DMM