beautypg.com

Specifications, S pecifications – Time Electronics 1030 Voltage & Current Source User Manual

Page 4

background image

4

1030 Technical Manual

P a g e

| 4

2.

S pecifications


Voltage Ranges:

0 - 1V (1mV resolution).

0 - 100 mV (100µV resolution).

0 - 10 mV (10µV resolution).

8 Volts (10mV resolution)

using external precision 1K

Ω resistor supplied.


Current Ranges:

0 - 100mA (100µA resolution),

0 - 10mA (10µA resolution).

Accuracy:
1 Volt range:

± 0.1% of FS ± 30µV.

100mV range:

± 0.1% of FS ± 3µV.

10mV range:

± 0.2% of FS ± 0.3µV.

100mA range:

± 0.2% of FS ± 3µA.

10mA range:

± 0.2% of FS ± 0.3µA.

8 Volt range:

± 0.3% of FS.


Linearity:

0.15% of setting.


Temp. Coeff:

150 ppm of FS /ºC (Outside 18 to 28ºC).


Noise:

30 ppm of full scale.


Battery:

9V PP3 type. Approx. 60 hours life

depending on output current.

Nicad rechargeable available as an optional extra.


Battery Condition:

Continuously monitored by front panel indicator.


Output Polarity:

Positive or Negative, switch selectable.

A centre Off position is also provided, which shorts

the output terminals together.

Maximum Output Current:

1V, 100mV Ranges:

Typically 20 mA.

10mV Range:

Up to short circuit value although it should be

noted that loads of less than 1k ohm will give

greater than 0.1% error.


Maximum Output Voltage
(Current Ranges):

8V.


Output Protection:

The 1030 can withstand continuous short circuit

or open circuit on all ranges.