8.
Maintenance and R epair.
8.1. Dismantling the Instrument
Remove rubber protection boot and then removal of four 6BA screws enables the
cover to be taken off which provides access to all parts of the instrument.
8.2. Battery Replacement.
Remove protection cover and then removal of four screws enables the cover to be
taken off which provides access to the battery lift battery out of holder and careful
disconnect battery connector, connect new battery and replace into case replace case
cover and recheck battery level.
8.3. Repair
NOTE: No repair work should be undertaken by the customer while the
instrument is under warranty as such work may render the warranty invalid.
Certain of the precision components used in this instrument are not readily available
and make repairs by the customer difficult if these components are damaged.
Overload conditions can cause a unit failure which will be indicated by one of the
following conditions:
a) Instrument inoperative and battery level indicator at zero.
b) Battery level indicator displaying but no output at the output terminals.