Electrical Data and Connections
Conductivity Input
Screw Terminal (3.5mm pitch)
Connection terminal
8-pin, 2-pin & 5-pin terminal blocks, DC
power connector (1.2mm diameter)
EMC Specifications
Emission
EN 61 36
Susceptibility
EN 61326
Other
Power Input
Universal +9 V DC Adaptor and 24V DC
Supply
Dimensions (W x H x D)
96mm x 96mm x 66 mm
Weight (Estimated)
210g
Ambient Temp. operating range
0 to 40
o
C
Maximum Relative Humidity
80% up to 31
o
C decreasing linearly to
50% at 40
o
C
Calibration Default Settings (CAL dEF)
Conductivity
Calibration factor
Reset to 1.000
Factory Default Settings (FCt dEF)
Conductivity
Calibration factor
Reset to 1.000
Cell constant
1.00
Cell type
2-cell
Measuring range
0 to 2000 µS/cm
Line resistance
0.000
Ω
Temperature
Unit of measurement
o
C
Temperature compensation
ATC On
Temperature coefficient
2.1% (Both process & calibration temp.
coefficients)
Temperature coefficient type
Linear
MTC (ATC Off)
25
o
C (Both process & calibration
temperature)
Controller Output
Controller type
Limit control
Relay status
De-energized
Relay 1 (SP1)
Set point value
100 µS
Relay function
Low