Thermo Fisher Scientific Alpha COND 560 User Manual
Page 37
![background image](/manuals/370691/37/background.png)
31
6
T E C H N I CAL S P E CI F I C AT I O NS
General Specification
Conductivity
Measuring Range:
Resolution:
0.000 - 2.000 µS/cm
0.001 µS/cm
0.00 - 20.00 µS/cm
0.01 µS/cm
0.0 - 200.0 µS/cm
0.1 µS/cm
0- 2000 µS/cm
1 µS/cm
0.00 - 20.00 mS/cm
0.01 mS/cm
0.0 - 200.0 mS/cm
0.1 mS/cm
Relative accuracy
± 1% of full scale reading
System accuracy
The effective measuring range and the
total system accuracy depend on the
sensor used with the controller. Please
refer respective sensor documentation.
Temperature
Measuring range
-10.0 to +125.0
o
C
+14.0 to +257.0
o
F
Resolution 0.1
o
C / 0.1
o
F
Relative accuracy
±0.5
o
C / ±1.0
o
F
Sensor
2 wire / 3 wire - Pt100 or Pt1000 (Jumper
Selectable)
Compensation
Automatic or Manual (reference at 25
o
C)
Calibration
Conductivity
Number of calibration points
1 point
Calibration window
±40% of default reading
Calibration standard solution
window
10% to 100% of the selected measuring
range
Temperature
Offset Adjustment
± 10 ˚C/ ± 18 ˚F
Controller
Output 2
SPST
Relays
Relay 1/Relay 2 function (for
Set Point 1 (SP1) & Set
Point(SP2))
LO (Relay is active when measured value
< Set point)
HI (Relay is active when measured value >
Set point)
Hysteresis window
0% to 10 % of selected measuring range
On-delay & Off-delay time
0 to 2000 seconds
Display
LCD
7 segments display with symbols for status
information
Back light
On/Off selectable