beautypg.com

Thermo Fisher Scientific Ion Selective Electrodes Sodium User Manual

Page 16

background image

Sodium Electrode

Instruction Manual

14

SPECIFICATIONS

Concentration Range:

Saturated to 1.0 x 10-

6

M (0.02 ppm)


pH Range:

5 to 12 (depending on Na

+

level)


Temperature Range:

-5

o

to 70

o

C


Resistance:

< 200 Mohm


Reproducibility:

±2%


Size:

110 mm length; 12 mm diameter; 1 m cable length


Storage:

Store in 5M NaCl with added ISA



ORDERING INFORMATION

Code No.

Description


EC-NA02

Sodium Ion Combination Electrode, epoxy body


EC-SCS-LNA1-BT

Sodium Standard, 0.1M NaCl


EC-SCS-LNA2-BT

Sodium Standard, 1,000 ppm Na

+


EC-SCS-LNA3-BT

Sodium Standard, 100 ppm Na

+


EC-ISA-LNA1-BT

Sodium Ionic Strength Adjuster, 4M NH

4

Cl/4M NH

4

OH