Instruction Manual
Sodium Electrode
13
TROUBLESHOOTING HINTS
Symptom
Possible Causes
Next Step
Out of Range Reading
defective meter
check meter with shorting strap
(see meter instruction manual)
defective electrode
check electrode operation
electrodes not plugged in
properly
unplug electrodes and reset
reference electrode not filled
replenish reference filling solution
electrodes not in solution
put electrodes in solution
air bubble on redipping
electrode
remove bubble by membrane
"Incorrect Answer" (but
calibration curve is
good)
incorrect scaling of semi-log
paper
plot millivolts on the linear axis. On
the log axis, be sure concentration
numbers within each decade are
increasing with increasing
concentration
incorrect sign
be sure to note sign of millivolt
number correctly
incorrect standards
prepare fresh standards
wrong units used
apply correct conversion factor:
1.0X10-
3
M = 23.0 ppm as Na
+1
sample pH too acidic
adjust pH with ISA
Drift (reading changing
in one direction)
samples and standards at
different temperature
allow solutions to come to room
temperature before measurement
incorrect reference filling
solution
use recommended filling solution
electrode exposed to
interference
soak overnight in electrode storage
solution
pH too acidic
adjust pH with ISA
ISA not used
use recommended ISA
Low Slope or No Slope
standards contaminated or
incorrectly made
prepare fresh standards
standard used as ISA
use ISA
pH too acidic
adjust pH with ISA
electrode exposed to
interference
soak overnight in electrode storage
solution
air bubble on membrane
remove bubble by redipping probe
Noisy or Unstable
Readings (reading
changing randomly)
defective meter
check meter with shorting strap
ISA not used
use recommended ISA
meter or stirrer not grounded
ground meter or stirrer
defective electrode
replace electrode
electrode exposed to
interference
soak overnight in electrode storage
solution