Instruction Manual

Ammonium Electrode

13

TROUBLESHOOTING HINTS

Symptom

Possible Causes

Next Step

Out of Range

defective meter

check meter with Reading

shorting

strap

(see meter instruction manual)


electrode(s)

unplug

electrode(s)

not plugged in

and reset

properly

reference electrode

be sure reference

not filled

electrode is filled


air bubbles on

remove bubble by

membrane

redipping

electrode


electrodes not

put electrode(s)

in solution

in solution



"Incorrect

incorrect scaling

plot millivolts on the linear

Answer"

of semi-log paper

axis. On the log axis, be sure

curve is good)

concentration numbers within

(but calibration

each decade are increasing

with increasing concentration


incorrect sign

be sure to note

sign of millivolt number correctly


incorrect standards

prepare fresh standards

wrong units used

apply correct conversion factor:

10-

3

M = 18 ppm NH

4

+ =

14 ppm as N


sample carryover

rinse electrodes

thoroughly between samples



Drift (reading

samples and standards

allow solutions

slowly changing

at different

to come to room

in one direction)

temperatures

temperature

before

measurement


electrode exposed

soak electrode

to interference

in ammonium standard


incorrect reference

use recommended

filling solution

filling solution