Thermo Fisher Scientific Ion Selective Electrodes Ammonium User Manual

Page 14

background image

Ammonium Electrode

Instruction Manual

8

Li

+1

2.0X10

1

2.0

2.0X10-

1


Na

+1

2.0X10

1

2.0

2.0X10-

1


+Tris

+1

is the cation of tris(hydroxymethyl)aminomethane


Interference

(ppm)

100 ppm NH

4

+

10 ppm NH

4

+

1 ppm NH

4

+


Cs

+1

1.0X10

2

1.0X10

1

1.0


K4

+1

2.7X10

2

2.7X101 2.7


Tl

+1

3.1X10

3

3.1X10

2

3.1X10

1


H

+1

1.6 pH

2.6 pH

3.6 pH


Ag

+1

2.7X10

5

2.7X10

4

2.7X10

3


+Tris

+1

3.1X10

5

3.1X10

4

3.1X10

3


Li

+1

3.5X10

4

3.5X10

3

3.5X10

2


Na

+1

1.1X10

5

1.1X10

4

1.1X10

3

Temperature Influences


Samples and standards should be at the same temperature, since electrode potentials are influenced
by changes in temperature. A 1oC difference in temperature results in a 2% error at the 1.0X10-

3

M

level.

Provided that temperature equilibrium has occurred, the ammonium ion electrodes can be used at
temperatures from 0o-50oC continuously and 40o-50oC intermittently. Room temperature
measurements are recommended, since measurements at temperatures markedly different from
room temperature may require equilibrium times up to one hour. Table 4 indicates the variation of
theoretical slope with temperature.

TABLE 4: Temperature vs Value for the Electrode Slope

Temp

(oC)

S"(slope)

0

54.20

10

56.18

20

58.16

25

59.16

30

60.15

40

62.13

50

64.11