beautypg.com

Thermo Fisher Scientific Ion Selective Electrodes Ammonium User Manual

Page 10

background image

Ammonium Electrode

Instruction Manual

4

TABLE

1: Concentration Unit Conversion Factors


ppm NH

4

+

ppm N moles/liter NH

4

+

1.80

1.40

1x10

-4

18.00

14.00

1x10

-3

180.00

140.00

1x10

-2



Measurement Procedure

Direct Measurement


A simple procedure for measuring a large number of samples. A single meter reading is all that is
required for each sample. The ionic strength of samples and standards should be made the same by
adjustment with ISA for all ammonium samples. The temperature of both sample solution and of
standard solutions should be the same.

Direct Measurement of Ammonium (using a standard pH/mV meter)

1. Prepare

10-

2

, 10-

3

, and 10-

4

M or 100, 10, and 1 ppm standards by serial dilution of the

0.1M or 1,000 ppm standard. Add 2 ml of ISA per 100 ml of standard.


2.

Place the most dilute solution (10-

4

M or 1 ppm) on the magnetic stirrer and begin stirring

at a constant rate. After assuring that the meter is in the mV mode, lower the electrode
tip(s) into the solution. When the reading has stabilized, record the mV reading.


3.

Place the midrange solution (10-

3

M or 10 ppm) on the magnetic stirrer and begin stirring.

After rinsing the electrode(s) with distilled water and blotting dry, immerse the electrode
tip(s) in the solution. When the reading has stabilized, record the mV reading.


4.

Place the most concentrated solution (10-

2

M or 100 ppm) on the magnetic stirrer and begin

stirring. After rinsing the electrode(s) with distilled water and blotting dry, immerse the
electrode tip(s) in the solution. When the reading has stabilized, record the mV reading.


5.

Using the semi-logarithmic graph paper, plot the mV reading (linear axis) against the
concentration (log axis). A typical calibration curve can be found in Figure 1.