beautypg.com

Thermo Fisher Scientific Ion Selective Electrodes Ammonium User Manual

Page 12

background image

Ammonium Electrode

Instruction Manual

6


11. Immerse the electrode tip(s) in the solution and wait for the reading to stabilize. Read the

concentration directly from the meter display.


12. The electrodes should be re-calibrated every 1-2 hours. Simply repeat Steps 2-8 (2-9)

above.

Low Level Ammonium Determination (using a standard pH/mV meter)


This procedure is recommended for solutions with ionic strengths less than 1.0 x 10-

2

M. If the

solution is high in ionic strength, but low in ammonium, use the same procedure, but prepare a
calibration solution with a composition similar to the sample.

1.

Using 20 ml of stock ISA, dilute to 100 ml with distilled water.


2.

Dilute 1 ml of the 0.1M standard to 100 ml to prepare a 1.0x10-

3

M standard solution for

measurements in moles per liter. Dilute 10 ml of the 1,000 ppm standard to 100 ml to
prepare a 100 ppm standard solution for measurements in ppm.


3.

To a 150 ml beaker, add 100 ml of distilled water and 1 ml of low level ISA. Place the
beaker on the magnetic stirrer and begin stirring at a constant rate.


4.

Place the electrode tip(s) in the solution. Assure that the meter is in the mV mode.


5.

Add increments of the 1.0x10-

3

M or 100 ppm standard as given in Table 2 below.


6.

After the reading has stabilized, record the mV reading after each addition.

TABLE 2: Stepwise Calibration for Low Level Ammonium Measurements

Added Concentration

Step

Pipet Volume (ml) M

ppm

1

A

0.1

1.0X10-

6

0.1

2

A

0.1

2.0X10-

6

0.2

3

A

0.2

4.0X10-

6

0.4

4

A

0.2

6.0X10-

6

0.6

5

A

0.4

9.9X10-

6

1.0

6

B

2

2.9X10-

5

2.9

7

B

2

4.8X10-

5

4.8

Pipet A = 1 ml graduated pipet

Pipet B = 2 ml pipet

7. On semi-logarithmic graph paper, plot the mV reading (linear axis) against the

concentration (log axis) as in Figure 1.


8.

Rinse the electrode(s) in distilled water and blot dry.