beautypg.com

Thermo Fisher Scientific CyberComm Pro for CyberScan pH 1100 & pH 2100 User Manual

Page 19

background image

Instruction Manual

CyberComm pH 1100/ 2100

15

Icon 14. Meter Settings

See Figure 19.
Program your meter functions via this icon on CyberComm Program.

a) Date and Time

Click on the textbox then a calendar will appear for
you to set the date.


b) Backlighting

Enable or disable the backlight and choose the On
period from 0 to 10 minutes.


c) Data

Log

Select Manual for self recording of the readings at
different times or choose to have Auto Data logging at
the preferred interval timing.


d) Memory

Recall

Select Manual or Auto.


e) Memory

Clear

Select Yes or No.


f) Data Transfer Mode

Select Current or Memory.


g) Reset all meter settings to meter default

Check box & click apply to reset Meter calibration settings.


h) Good Laboratory Practice Settings

Enter the relevant information for useful future references.


Click on OK or Apply to save settings whenever there are changes.



Continue from Section 4.3

Refer back to Figure 12 on pg 13.

4.3.1 Measurement

Mode

This indicates the current mode of measurement of the meter, you can change the mode by selecting from
the drop down menu to choose between the various parameters; pH, mV and Ion. (Ion mode: available in pH
2100 only)


4.3.2 Show

Cursor

Use the ‘Show cursor’ function from the icon panel and a green guide line will show any point on the reading
curve having the exact measurement according to the mode chosen. Drag this guide along the horizontal
time axis and you can read the pH, mV or concentration values at any point in time.


4.3.3 Information

This information will always be on the graph, showing the Elapsed Time of the measurement being carried
out, the date and time which marks the start of the first data captured.

4.3.4 Data

Reading

The data captured from the meter is plotted on the graph while the measurement is being carried out.

Figure 19: Meter Settings