Thermo Fisher Scientific CyberComm Pro for CyberScan pH 1100 & pH 2100 User Manual

Page 16

background image

Instruction Manual

CyberComm pH 1100/ 2100

12

Stability definitions

High stability:
reading is stabilized quickly, but reading is less accurate.
Medium stability:
reading stability is averaged between high and low stability.
Low stability:
reading is stabilized slowly, but guarantees high accuracy

Icon 11. pH Settings

See Figure 16.


a) pH

Resolution

Select the accuracy of your readings by selecting either 0.001 pH
resolution or 0.01 pH resolution.


b) Temperature

Scale

Select °C or °F.


c) pH Alarm Limits

Using the Low and High options, select the ‘safe’ range of
readings. Any readings not within the selected range will trigger
the alarm alert.


d) pH Calibration Due

Check to enable the calibration due warning or uncheck to disable
it. Calibration due period can be set from 1 to 365 days to remind
you of the calibration due


e) pH

Stability

Check the box to enable pH Stability Indicator; indicator is
disabled when box is unchecked. Select from Low, Medium to
High to get the preferred level of stability in readings respectively.
(Low for low stability in readings, Medium for medium stability

and High for high stability in readings)


f) Reset pH Calibration Setting

Check box & click apply to reset pH calibration settings


Click on OK or Apply to save settings whenever there are changes.

Figure 16: pH Settings