Thermo Fisher Scientific EcoScan pH/Ion 5 & 6 User Manual

Page 19

background image

Instruction Manual

pH 5/6 & Ion 5/6

- 16

-

3.5.2

Without Temperature probe (no ATC)

If no temperature probe is used, the meter compensates for pH response based on a
new calibrated temperature value manually set by you or at 25.0

o

C (factory default).

1. Press

MODE key to enter into Temperature mode until “

o

C” shows in LCD.

2. Compare the displayed value to NIST certified thermometer or thermometer known

to be accurate (dipped into a constant temperature bath).

3. Press CAL key to enter temperature calibration mode. The LCD shows “CA

momentarily and displayed reading flashes. Note that this displayed value should
either be 25.0

o

C or last set temperature value.

4. Press and

key (for Ion 6) or INC key (for pH 5/6) until the displays shows the

desired temperature. You can set any value from 0 to 100

o

C.

5. To cancel or abort this operation, press CAL key. Note no new value will be stored

into its meter‘s non-volatile memory. To confirm calibration, press ENTER key. The
LCD displays “CO” momentarily, and the meter reverts to measurement mode.