3.2.2

pH Calibration using USA standard buffers

1. Pour a known pH buffer standard solution into a clean, dry container, e.g. pH 7.00.

Power on the meter and it automatically enters into measurement mode. Select pH
mode by pressing MODE key if necessary.

2. Dip both pH electrode and temperature probe into pH 7.00 buffer solution. Swirl

gently and wait for reading to stabilise (approx. 30 seconds depending on your
electrode condition).

3. Press

CAL key to enter pH calibration mode. A “CA” displays momentarily and the

display shows the current uncalibrated reading flashing while in the calibration
mode.

4. To abort or cancel calibration without accepting new value, press CAL key. The

meter then reverts to pH measurement mode.

5. To proceed calibration, allow reading to stabilise. The meter automatically

recognises pH 4.01, 7.00 or 10.01 buffers. Press ENTER key to confirm calibration
and the LCD displays “CO” momentarily. The meter reverts to measurement mode.

6. For 2 or 3-point calibration, repeat step 3 with other pH buffer values of 4.01 and/or

10.01 for higher accuracy.