Thermo Fisher Scientific EcoScan pH/Ion 5 & 6 User Manual

Page 15

background image

Instruction Manual

pH 5/6 & Ion 5/6

- 12

-

3.3.1 Multi-point

Ion

Calibration


1. In case if you need to start calibration using 1.0 ppm standard solution, you can skip

the 0.1 ppm calibration option (as mentioned * below).

2. For this instance let’s take 1.0 ppm calibration. To begin, pour a known 1.0 ppm

standard solution into a clean container. Power on the meter, and set the meter to
Ion measurement mode.

3.

Dip the ISE into 1.0 ppm standard solution. Swirl it gently. Press CAL key to enter
into calibration mode.

4.

The display shows “CA” momentarily followed by “0.1” flashing.

* At this moment, if you wish to calibrate 1

st

ion calibration option other than 0.1 ppm,

use

key once to select 1.0 ppm. Using

or key you may select other options

like 0.1, 10.0 or 100.0 ppm.

5. Press ENTER key and the displayed value shows corresponding measured mV

reading that is equivalent to 1.0 ppm. Allow the reading to stabilise.

6. Press

ENTER key to confirm 1

st

point calibration (e.g. 1.0 ppm) and the display will

show the next calibration option, 10.0 ppm. Rinse the electrode with distilled water
and blot it dry if necessary.

7. Pour a known 10.0 ppm standard solution into another clean container. Dip the

electrode into the standard solution. Swirl it gently.

8. Press

ENTER key and the displayed value shows measured mV reading that is

equivalent to 10.0 ppm. Allow the reading to stabilise.