Instruction Manual

pH 5+, pH 6+, Ion 6+

14

Without Temperature Probe (Manual Temperature Compensation)

If a temperature probe is not connected, the meter compensates for pH response based
on the factory default value at 25.0 °C. This default temperature can be manually
adjusted using the procedure below.

For nearly all applications however, Automatic Temperature Compensation (ATC) is
recommended for best accuracy.

1.

Turn on the meter and select °C mode by pressing

if necessary. Disconnect

the temperature probe.

2. Press

to begin temperature calibration mode. “

” (calibrate) will display

briefly. The factory default temperature (25.0 °C value will blink.

3. Use

(pH 5+ and pH 6+) or

(Ion 6+) to adjust the reading to the

desired value. The maximum adjustment is ± 0 to 100 °C.

4. Press

to confirm the value. “

” (confirm) will display briefly. The meter will

automatically save the calibration, cease blinking, and begin °C measurement with
the new default temperature.