beautypg.com

Thermo Fisher Scientific Eutech pH 5/6 Plus & Ion 6 Plus (New version R1.1, SN >797406) User Manual

Page 13

background image

Instruction Manual

pH 5+, pH 6+, Ion 6+

11

5. When the mV reading has stabilized, the mV annunciator will stop blinking. Press

to confirm the value. “

” (confirm) will display briefly. The display will show

the next highest calibration standard value. Rinse the electrode with clean water.

6. Repeat steps 3 & 4 & 5 once for a 2-point calibration or twice for a 3-point

calibration using additional ion calibration standard(s).

7. Press

to complete a 2-point calibration. When a 3-point calibration has been

performed, the meter automatically completes the calibration.

8. A successful calibration will show “

” (slope) followed by the mV/decade value,

the display will cease blinking and begin ion measurement.

Error message “



” is displayed if

is pressed after only one point calibration has

been completed. Recalibrate using minimum of 2 points.

Error message “



” is displayed if

is pressed when the completed calibration

points are not consecutive. Recalibrate with calibration standards that are 1 decade
apart from each other.

Error message “



” is displayed if the calibration is not successfully stored into

memory. This occurs when the slope is <15 mV/decade or >90 mV/decade.