Instruction Manual

pH 5+, pH 6+, Ion 6+

9

Resetting User Calibrated Values

The meter can be completely reset to factory default values, or partially reset for
calibration values only using the procedure below.

Reset Type

Description

No (

)

No reset is performed; meter returns to measurement

Calibration (

)

Reset of either pH, mV, Ion, or Temperature calibration

depending on the selected mode*

Factory (

)

Reset of all calibration values and user settings to factory

default settings

*For calibration reset, measure in the mode that you want to reset prior to step 1 below.

1.

Power off the meter. Press and hold

then press

. If successful,”



” will

blink on the display. Release both keys.

2. Press

to enter the reset menu.

3. Press

to toggle between No Reset (

), Calibration Reset (), or Factory

Reset (

). Press

to cancel.

4. Press

to confirm the selected reset type. The meter will automatically begin

measurement mode.