Techne TC-3000G User Manual
Page 30

[
to disable ]
[
turn on/off ]
28
To edit a temperature, type in the required
temperature. In the example here, the heated lid
temperature is being edited.
Press "Enter" to accept.
Use the "Down Arrow" and/or the "Up Arrow"
keys to select the next field or step you
wish to edit.
To edit a field with on/off settings, first select the
field and press "Enter"
Use the "Up Arrow" and "Down Arrow" keys to
toggle between the settings.
Press "Enter" to accept.
Sample cooling enables the block to be actively
cooled to 4°C whilst the heated lid is reaching
the correct temperature.
To edit a step with both a temperature and a
time, first press "Enter" and type in the new
temperature. Press "Enter" again. Now type
in the new time. The default is minutes.
1. For whole minutes, type the number required
then "Enter" .
2. To toggle between minutes and hours, use the
"Decimal Point" key.
3. For values less than 1 minute, first press the
"Decimal Point" key followed by the number
of seconds required.
4. For minutes and seconds, type the number of
minutes followed by the "Decimal Point" key
then the number of seconds.
Once the values have been added, press "Enter"
to accept.
Once you have finished editing and you are
happy with the changes, press the "End" key
followed by the "Enter" key to save.
If you press the "End" key a second time for
'No' then all the changes you have made will be
lost.
DONNA
Heated Lid
105°C
Preheat lid
on
[ edit temperature ]
DONNA
Heated Lid
105°C
Preheat lid
on
[
to change ]
[
to disable ]
[ .for hrs/mins ]
Pause
off
-------------------------
In denat 4.0°C
00 m30
[ edit temperature]
9
Pause
off
-------------------------
In denat 94.0°C
0 m30
[ edit time ]
0
-------------------------
Fin extn 72.0°C
5m00
Fin hold
10.0°C
[ , to edit ]
[ to save ]
DONNA
Preheat Lid
on
Sample cooling
off