Heated Lid

Use this to set the heated lid temperature (in general approximately

10ºC higher than the hottest block temperature in a program).

Preheat Lid

Set to ON to pre-heat the lid to the set temperature before the thermal

cycling program begins.

Sample Cooling

Set to ON to cool the block to 4ºC while the lid preheats.

Pause

Set to ON to pause at the ambient temperature of the block before the

program starts, unless Sample Cooling has been set to ON. In this case

the unit will pause with the block held at 4ºC. The program will not

commence until the unit is manually un-paused.

Initial denaturation

Set to ON for the required hold time to denature the sample and/or

activate the enzyme before the thermal cycling program begins.

Hot Start

Set to ON to pause at the required set temperature. The program will

not continue until the unit is manually un-paused.

Final Extension

Set to ON for the required hold time to add a final extension step to the

program to ensure complete amplification of the products.

Final Hold

Set to ON to hold the block at the required set temperature at the end of

a program until the user manually stops the unit.

Pause

Set to ON to pause the unit before the program starts. The block will

pause at its current temperature unless Sample Cooling has been set to

ON. In this case the unit will pause with the block held at 4ºC and the

program will not commence until the unit is manually un-paused.

Thermal Cycling

Enter appropriate values in the parameter boxes to create the thermal cycling program. Note: if

Display Invalid Data Warnings has been set to ON in the settings, any invalid values will cause the
box to glow yellow and a warning will be displayed at the bottom of the screen advising on the correct

range.

Stage name

Type in a stage name if required. Up to 20 characters can be used.

Cycles

Enter the number of cycles (repeats) for the stage. The maximum is 99.

Ramp rate

Select the required ramp rate. The range is 0.1ºC/s (minimum) to

5.0ºC/s maximum for heating and 3.0ºC/s for cooling. The value can be

entered to 1 decimal place. Ramp rate will affect the heating or cooling

rate from the previous step to the current step.