beautypg.com

Techne RS232 User Manual

Page 25

background image

D B - 3 D w i t h R S 2 3 2 O P E R A T O R ’ S M A N U A L

P A G E 23

After Use

1. When you have finished heating samples, remember that parts of the unit –

the tubes, blocks and associated accessories – may be very hot. Take the
precautions listed earlier. We recommend that the blocks should be allowed
to cool to 70°C before being removed from the Dri-Block

®

. They will still

have to be handled with care

2. If you need to remove an insert block while it is hot, screw the extractor tool

into the thermometer hole and lift the block out vertically. Never leave the
extractor tool in the block while it is being used in the Dri-Block

®

.

Removing an insert block with the extractor tool

This manual is related to the following products: