beautypg.com

Specification – Techne RS232 User Manual

Page 17

background image

D B - 3 D w i t h R S 2 3 2 O P E R A T O R ’ S M A N U A L

P A G E 15

SPECIFICATION

The DB-3D with RS232 Dri-Block

®

This unit has a recessed chamber which can hold up to three interchangeable
insert blocks which accept test tubes or other sample containers. The
Temperature is set by 3 front panel buttons with digital display.

The DB-3D is supplied with an extraction tool for removing the insert blocks
from the unit.

Working temperature range

Ambient +5°C to 200°C

Settable temperature range

0.0°C to 200.0°C

Temperature stability

±0.1°C at 40°C

Temperature stability

±0.15°C at 100°C

Temperature setting

3 push buttons

Temperature display

4 digit LED

Set point resolution

0.1°C

Set point to accuracy

±1°C

Max temperature variation between blocks

0.2°C at 40° C

Heat up time 30° C to 37° C

11 minutes

Heat up time 30° C to 56° C

15 minutes

Heat up time 30°C to 200°C

25 minutes

Number of blocks

3

Electrical supply
Voltage

Cycles

Power

230V

50Hz-60Hz

650W

110V-120V

50Hz-60Hz

650W

This manual is related to the following products: