Multi-parameter transmitter user’s manual, 6 paddle wheel flow sensor connections and setup – Metex SharkTX Multi-Parameter Transmitters User Manual
Page 12

Section 3 - Electrical Connections and Setup
Page 10
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.6 Paddle Wheel Flow Sensor connections and setup
Figure 3.5 Connections for Flow Sensor
SETUP OF UNITS
OF VOLUME
SETUP OF UNITS
CALIBRATION FACTOR
ENTER FLOW SENSOR
SELECT FLOW METER
SEC. 7.6
OF TIME
SEC 7.5
SEC. 7.4
SEC. 7.1
TOTAL 0
TOTALIZER RESET
UNITS OF TIME
UNITS OF VOLUME
K FACTOR
METER SELECTION
RESET TOTALIZER
RUN MODE
TO ZERO
SEC 7.0
Once connected, step through
the LCD menus to select the
sensor in the order shown. The
Sensor K factor (pulses per U.S.
Gallon) is usually printed on the
side of the sensor or on a label
attached to the sensor cable.
The drawing shows the connnections for a typical paddle wheel flow sensor. The cable to the
sensor should not exceed 2000’ (600 meters).
Dwg# N104-45