Millivolts, Tc (thermocouples) – Martel Electronics 235 User Manual
Page 51

achieved at all times. The internal microprocessor will automatically
correct for any drift due to temperature or battery change.
Millivolts
The millivolts output function is identical to the volt function except
that the value is entered and displayed in millivolts.
TC (Thermocouples)
The Model 230 calibrator supports 11 thermocouple curves,
JKTERSBNCGD.
The Model 235 calibrator supports 15 thermocouple curves,
JKTERSBNCGDLtMP. The additional curves on the Model 235 are:
L = J Thermocouple (Din Curve)
t = T Thermocouple (Din Curve)
M = Ni-Ni Moly
P = Platinel
Both calibrators can read out in either degrees Celsius or Fahrenheit.
To simulate a thermocouple output, press the TC key when the
calibrator is in the output mode. The calibrator will display a prompting
menu asking for the type of thermocouple.
Model 230 Display:
Model 230 (Optional):
Model 235 Display:
Output Functions
230/235
46
Output TC type
JKTERSBNCGD
Select Temp Std
1948 1968
Output TC type
JKTERSBNCGDLtMP
Select Temp Std
1948 1968 1990