7. Setpoint Feature

The TC100 allows the user to program up to 3
setpoints for each TC type (including mV). to
set the setpoints complete the following proce-
dure:

1) Turn on the TC100 and select the Source

Mode.

2) Select the TC type to be simulated and cor-

rect units (either °C or °F).

3) Press

to see what is currently

stored in each of the 3 setpoints.

4) To store a new value adjust the output to the

desired value then press the

key.

5) Repeat steps 3 & 4 until all 3 setpoints are

stored.

6) Once the setpoints are stored, press the

key to recall the newly stored set-

points.

7) Each press of the

key will incre-

ment to the next stored setpoint.

Note: Setpoints are non-volatile and will
remain in memory until changed.

8. Min/Max Datalogging

When the TC100 is in the measure mode it will
continuously monitor and store the minimum
and maximum values for that particular input.
To operate the datalogging feature use the fol-
lowing procedure:

Recall

Recall

Set

Recall