Martel Electronics M2000 User Manual
Page 23
23
*OPT?
IEEE-488
RS-232
Sequential
Overlapped
(Options command) Returns a list of the installed hardware and software options.
Responses:
0 (no options are installed)
Example:
*OPT? Returns 0
This command is reserved for future use.
OUT
IEEE-488
RS-232
Sequential
Overlapped
(Output command) Sets the output of the M2000 and establishes a new reference point for
the error mode. To source or measure a temperature, select the desired sensor and sensor
parameters first. (See the TSENS_TYPE, RTD_TYPE, and TC_TYPE commands.)
Use multipliers, k, m, u with the OUT command, as desired.
Parameters:
Volts dc
Current dc
Examples:
OUT 15.2 V
(volts; 15.2 V)
OUT 1.2 mA
(current; 1.2 mA)
Each example shows a value and unit, e.g., -15.2 V. If a value is entered without a unit, the
value of the existing output is changed, when logically allowed.
OUT?
IEEE-488
RS-232
Sequential
Overlapped
(Output query) Returns the output amplitude of the M2000. Multipliers (e.g., K or M) are not
used in the response.
Response: value>, Examples: OUT? returns -1.520000E+01,V The respective values for the above examples are: -15.2 V RANGE? IEEE-488 RS-232 Sequential Overlapped (Range query) Returns the present voltage output range. Response: Examples: V_0.1V (dc volts 100 mV range) V_1V (dc volts 1V range) V_10V (dc volts 10 V range) V_100V (dc volts 100 V range) Returns the symbolic name of the voltage output range. RANGELCK IEEE-488 RS-232 Sequential Overlapped (Range lock command) Locks or unlocks the present voltage range. Parameter: ON Locks the present voltage range OFF Unlocks the present voltage range. RANGELCK? IEEE-488 RS-232 Sequential Overlapped (Range command) Returns the RANGELOCK state 1 for ON or 0 for OFF. Parameter: (None) X X X X X X X X X X X X X X X X X X
OUT? returns 1.88300E-02,A
OUT? returns 1.23000E+00,V
OUT? returns 1.52000E+01,V
18.83 mA
1.23 V
15.2 V