Error code listing, Rs-232 port commands, Status commands – Martel Electronics M2000 User Manual
Page 20: Remote command listing
![background image](/manuals/369781/20/background.png)
RS-232 Port Commands
Command Description
LOCAL
Puts the M2000 into the local state and disables lockout.
LOCKOUT
Puts the M2000 into the lockout state. This command duplicates the
IEEE-488 LLO
(Local Lockout) message.
REMOTE
Puts the M2000 into the remote state. This command duplicates the
IEEE-488 REN
(Remote Enable) message.
Status Commands
Command Description
FAULT?
Returns the most recent error code in the M2000 error queue, then removes that error
code from the queue.
Error Code Listing
Error
Message
Number
Class Description
1
DDE
Error queue overflow.
101
CME
A non-numeric entry was received in a field that should contain a numeric
entry.
102
EXE
The numeric field exceeds 10 characters.
105
EXE
Entry is above upper limit for the selected output range.
106
EXE
Entry is below lower limit for the selected output range.
107
EXE
OUT? was received when the M2000 was not in a Source mode.
108
CME
A required command parameter was missing.
110
CME
An invalid RANGELCK parameter (not ON or OFF) was received.
111
EXE
RANGELCK ON was received when the M2000 is not in Volts mode.
117
CME
An unrecognizable command was received.
118
CME
An invalid parameter was received.
120
EXE
The serial input buffer overflowed
121
EXE
The command string buffer overflowed.
122
QYE
The serial output buffer overflowed.
123
DDE
The output overloaded.
124
DDE
M2000 is out of tolerance. This error is set after a failed initialization or a
failed *TST?.
125
DDE
M2000 has a ADC failure. This error is set after a failed initialization or a
failed *TST?.
Remote Command Listing
The following is an alphabetical list of all M2000 commands and queries, including common
commands and device-dependent commands. Each command title includes a graphic that
indicates remote interface applicability, IEEE-488 and RS-232, and command group:
Sequential and Overlapped.
*CLS
IEEE-488
RS-232
Sequential
Overlapped
(Clear Status command) Clears the ESR, the error queue, and the RQS bit in the status byte.
This command terminates pending operation complete commands (*OPC or *OPC?).
Parameter: (None)
Example: *CLS
Clear the ESR, the error queue, and the RQS bit in the status byte.
X
X
X