2 measuring pressure with external modules, 3 measuring and sourcing current (4 ma), Pressure module adapter pressure module valve – Martel Electronics DPC300A User Manual
Page 18

16
16
4.2 Measuring pressure with external modules
The calibrator has a digital interface for external pressure modules.
These modules are available for various ranges, including gauge,
vacuum, differential and absolute pressure. The modules work
seamlessly in conjunction with the calibrator. Simply plug them into
the interface and select [EXT] (external sensor). Since the interface
between the calibrator and the module is digital, the accuracy and
display resolution is dependent on the module.
REGULATOR
MAX 120 PSI
AIR SUPPLY
0-100 PSI
OUTPUT
-12-300 PSI
PUMP
-12-300 PSI
MEASURE
FINE ADJUST / VENT
PUMP
SELECTOR
MAIN
SELECTOR
WARNING:
VENT BEFORE SELECTION
HOME
PUMP
ZERO
J1
J2
J3
COM
VOLTS /mA
SWITCH TEST +24v DC
PRESSURE
MODULE
F1
F2
F3
FILTER
CHARGE
PRESSURE
VACUUM
P2
P3
P4
MARTEL ELECTRONICS
DPC-300A
P1
P2
P3
P4
Pressure Module
Adapter
Pressure Module
Valve
4.3 Measuring and sourcing current (4 ... 20 mA)
1. To measure current, use the input terminals on the front of the
calibrator. Select the mA function on the lower display. Current is
measured in mA and percentage of the measuring range. The
measuring range on the calibrator is set to 0 % at 4 mA and 100 %
at 20 mA.
Example
If the current measured is displayed as 75 %, then the value is 16
mA.
Note: The display will indicate "OL" when the measured
current exceeds the nominal range of current measurement
(24 mA).
2. The same connections are used for the current source. Select
mA-Source or mA Sim-2W from the configuration display.