To set loop power on send ERROR_LOOP ON.
ERROR _LOOP?
Returns the current state of loop power in percent error mode.
For example:
If you send ERROR_LOOP? It will return ON if loop power is on in error mode.
ERROR_ MODE
Turns percent error mode on and off.
For example:
To turn on percent error mode send ERROR_MODE ON.
ERROR _ MODE?
Returns the current state of percent error mode.
For example:
If you send ERROR_MODE? It will return ON if the calibrator is in percent error
mode.
ERROR_ PORT
Sets the pressure port for percent error.
For example:
To set the pressure port for percent error to [P1] send ERROR_ PORT P1.
ERROR _ PORT?
Returns the current pressure port for percent error mode.
For example:
If you send ERROR _PORT?, it will return P1 if the pressure port in percent
error is [P1].
FUNC
Sets the display indicated in argument one to the function indicated in argument 2.
For example:
To set the lower display to P1 Pressure mode send FUNC LOWER,P1.
FUNC?
Returns the current mode of all displays. For example if the calibrator is set to [P1]
ST on the upper display, [P1] on the middle, and the lower display is off, FUNC?
Would return: