Display
: 1999counts (3 1/2digits), Large LCD
●
Used location
: Altitude up to 2000m, indoor use
●
Operating temperature : 0ºC ~ 40ºC, relative humidity 85%
& humidity
(no condensation)
●
Storage temperature
: -20ºC ~ 60ºC, relative humidity 85%
& humidity
(no condensation)
●
Withstand Voltage
: AC3700V / 1 min
(between electrical circuit and enclosure)
●
Insulation resistance
: 50MΩ or more / 1000V
(between electrical circuit and enclosure)
●
Sleep Function
: 1. Automatically enters Sleep mode in 3 min
after the last switch operation (current
consumption
75uA).
This function doesn’t
work at voltage measurements. To exit
from the Sleep mode, set the Function
switch to OFF position once, and re-set it
to the Range at which a measurement to
be
conducted.
2. Backlight turns off in 1 min after it lights
up.
●
Dimension
: 186mm x 167mm x 89mm
●
Weight :
965g
●
Power Source
: DC12V / Size AA battery R6P(SUM-3)
x
8pcs
●
Possible number of
: 1200 times or more
Measurements
measure every 30sec at x1/2 Range,
IΔn =100mA)
●
Accessories
: Instruction manual x 1pce
Strap belt x 1pce
Test lead M7128 x 1set (red & black cords)
Test lead with alligator clip M7129 x 1 set
Cord case x 1pce
Long pin M8017 x 2pcs
Size AA battery R6P(SUM-3) x 8pcs
6