Specification – KYORITSU 5410 User Manual
Page 7

4. Specification
●
Measuring range and accuracy
(23ºC±5ºC, relative humidity 75% or less)
Range
Rated
Voltage
Test current
IΔn
Measuring
range
Accuracy
Trip Time
±(1%rdg+3dgt)
x 5
10 / 30 / 50
/ 100mA
Testing time
0ms ~ 200ms
Test Current
+2% ~ +8%
Trip Time
±(1%rdg+3dgt)
x 1
Testing time
0ms ~ 2000ms
Test Current
+2% ~ +8%
Trip Time
±(1%rdg+3dgt)
x 1 / 2
10 / 30 / 50
/ 100 / 200
/ 500mA
Testing time
0ms ~ 2000ms
Test Current
-8% ~ -2%
AUTO
RAMP
(mA)
100V±10%
200V+32%
/-10%
400V±10%
50 / 60Hz
10 / 30 / 50
/ 100 / 200
/ 500mA
40%~ 110% of IΔn
(goes up by 5%)
Testing time
300ms x 15steps
Test Current
at each step
-4% ~ +4%
* Only the RCD type G (without trip out time-delay) can be tested at
Auto Ramp Test; type S (time-delay) cannot be tested.
Voltage Measurement
Measuring range
Accuracy
80V ~ 450V
50 / 60Hz
±(2%rdg+4dgt)
●
Applied standards
: IEC61010-1 Measurement Category
CAT.III 300V / CAT.II 400V,
Pollution
degree
2
IEC61010-031
IEC61557-1,
6
IEC60529 IP54