beautypg.com

Specification – KYORITSU 4300 User Manual

Page 7

background image

̶ 5 ̶

3. Specification

● Measurement range and accuracy (23ºC±5Cº, relative humidity

75% or less)

Voltage/ Earth voltage measurement

Measurement range

Display range

Accuracy

AC5.0 to 300.0V

(45 to 65Hz)

(425V peak or less)

0.0 to 314.9V

±1%rdg±4dgt

DC±5.0 to ±300.0V

0.0 to ±314.9V

±1%rdg±8dgt

*AC measurement method : rms detection

For the waveforms other than sinewave

of CF<2.5, the specified accuracy

±1%f.s. should be applied.

* AC/DC auto-detection at 5V or higher

* DC sign appears at 5V or higher.

(H+S(C+P) terminal: positive side, E terminal : negative side)

* The LCD reads 0.0V for 0.9V or lower voltages.

Earth resistance measurement

Range

(auto-ranging)

Measurement

range

Display range

Accuracy

200Ω Range

5.0 to 2000Ω

0.0 to 209.9Ω

±3%rdg±5dgt

2000Ω Range

160 to 2099Ω

Measurement method: Constant current inverter

approx. 1.4mA (200Ω Range)/825Hz

approx. 0.7mA (2000Ω Range)/825Hz

Open voltage: Approx. 13V