Specification – KYORITSU 4300 User Manual
Page 7
̶ 5 ̶
3. Specification
● Measurement range and accuracy (23ºC±5Cº, relative humidity
75% or less)
Voltage/ Earth voltage measurement
Measurement range
Display range
Accuracy
AC5.0 to 300.0V
(45 to 65Hz)
(425V peak or less)
0.0 to 314.9V
±1%rdg±4dgt
DC±5.0 to ±300.0V
0.0 to ±314.9V
±1%rdg±8dgt
*AC measurement method : rms detection
For the waveforms other than sinewave
of CF<2.5, the specified accuracy
±1%f.s. should be applied.
* AC/DC auto-detection at 5V or higher
* DC sign appears at 5V or higher.
(H+S(C+P) terminal: positive side, E terminal : negative side)
* The LCD reads 0.0V for 0.9V or lower voltages.
Earth resistance measurement
Range
(auto-ranging)
Measurement
range
Display range
Accuracy
200Ω Range
5.0 to 2000Ω
0.0 to 209.9Ω
±3%rdg±5dgt
2000Ω Range
160 to 2099Ω
Measurement method: Constant current inverter
approx. 1.4mA (200Ω Range)/825Hz
approx. 0.7mA (2000Ω Range)/825Hz
Open voltage: Approx. 13V
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)