Voltage measurement, 1 measurement procedure – KYORITSU 4300 User Manual
Page 15
̶ 13 ̶
7. Voltage measurement
#
DANGER
● Never apply voltages exceeding 300V, max over-voltage
protection, to this instrument.
7-1 Measurement procedure
(1) Connect the test leads to the instrument.
Plug the L-shaped Banana plug of MODEL7248 into the E
terminal as illustrated below.
(2) Connect the adapter for test leads to the earth side of the circuit
under test and the metal tip (H+S(C+P) terminal) to the line side
respectively. Connecting that in reverse is also fine if the circuit
under test is not earthed.
Caution
Do not press
the Test button.
Fig. 7-1
Fig. 7-2