KYORITSU 3132A User Manual
Page 8
― 6 ―
Continuity Test (Resistance Test) Ranges(IEC 61557-4)
Ranges
Measuring range to keep
Maximum percentage
operating error
operating error
3
Ω
0.2
Ω
- 3
Ω
±30%
The influencing variations used for calculating the operating error are denoted
as follows:
Temperature
: 0
。
C and 35
。
C
Supply voltage : 6.4V to 10.4V
Position
: Reference position ±90
。
※Prior to measurement, apply 0-Adjustment at each position.
○ AC Voltage Warning:
Warning range
0-600V
Accuracy
±5% of scale length
Input impedance
1.2M
Ω
● Typical Number of measurements (central tendency for supply voltage up to
6.0V)
Insulation Resistance Ranges:
1M
Ω
at 1000V Range
Approx. 1,400 times min.
0.5M
Ω
at 500V Range
Approx. 3,500 times min.
0.25M
Ω
at 250V Range
Approx. 5,500 times min.
Continuity Test (Resistance Test) Ranges:
1
Ω
at 3
Ω
Range
Approx. 1,500 times min.
● Applicable Standards
IEC 61010-1
Overvoltage CAT.III 600V Pollution Degree 2
IEC 61010-2-31
Safety requirements for hand-held probe assemblies
IEC 61557-1/2/4
Measuring equipment for low voltage distribution systems
IEC 61326-1
EMC
IEC 60529 (IP54) Dust & drip proof
● Operating Temperature & Humidity:
0 - 40°C, relative humidity up to 85%
● Storage Temperature & Humidity:
-10 - 50°C, relative humidity up to 75%
● Insulation Resistance :
More than 50M
Ω
at 1000V DC between electrical circuit
and housing case
● Withstand Voltage: 5550V AC for one minute between electrical circuit and
housing case
● Overload Protection Insulation resistance ranges:
1000V Range
1200V (DC+AC p-p) for 10 seconds
500V Range
600V (DC+AC p-p) for 10 seconds
250V Range
300V (DC+AC p-p) for 10 seconds
Continuity ranges:
3
Ω
/500
Ω
Range 280V (DC+AC p-p) for 10 seconds
AC Voltage Warning:
1200V (DC+AC p-p) for 10 seconds
● Dimensions:
106(L) x 160(W) x 72(D) mm approx.
● Weight:
560g approx. (including batteries)
● Power Source:
6 x 1.5V battery type R6P ,1.5V AA or equivalent
● Accessories
Test Probe MODEL7122
x 1 set
Pouch for test Probe
x 1
Shoulder strap
x 1
R6P batteries
x 6
Spare fuse F500mA/600V x 1
Instruction manual
x 1
― 7 ―
3132A-E/03-08 03.8.6 10:33 AM ページ 6