KROHNE BM 702 EN User Manual
Page 13
Installation and operating instructions BM 702 (06/04)
13
Function (Fct.)
Input range
Description
(compensates different
wave speeds in still wells)
3.1.7 REF.OFFSET
Enter -10.00...0...+10.00 [m] Reference offset is added
to measured distance
values.
3.1.8 TB.OFFSET
Enter-100.00...0...+100.00 [m] Tank bottom offset is added
to measured level values.
3.2 DISPLAY
3.2.1 FCT.DISP
Select LEVEL
DISTANCE
CONVERSION
Select function of display
(value to be displayed).
See also explanatory notes.
3.2.2 UNIT.LENGTH
Select m/cm/mm/
inch/Ft/
PERCENT/BARGRAPH
Select unit for length value
to be displayed (only for
level and distance).
3.2.3 UNIT.CONV.
Select m3/l(Liter)/US Gal/
GB
Gal/Ft3/bbl/PERCENT/
BARGRAPH/USER
UNIT
Select unit for conversion
value to be displayed
(“volume table“).
(see explanatory notes)
3.2.4
USER UNIT
Text entry
10 characters
Enter user-defined unit for
the conversion table.
3.2.5 ERROR
MSG.
Select NO/YES Select whether error
messages to be shown in
display.
3.3 SIGNAL
OUT
3.3.1 FUNCTION
I
Select OFF/LEVEL/DISTANCE/
CONVERSION/SW.OUTP.
Select function of the
current output.
3.3.2
RANGE I
Select 3.8-20.5mA/E3.6
3.8-20.5mA/E22
4-20mA
4-20mA/E3.6
4-20mA/E22
Select range/error status
for the current output (hold
last value or 3.6 mA/22mA
in error status)
3.3.3
SCALE 4mA
Enter -200.00 ... +200.00 [m]
0.00 ... 99999.99 [m
3
]
Enter lower measuring
range value for the current
output (4 mA).
(see explanatory notes)
3.3.4
SCALE 20mA
Enter -200.00 ... +200.00 [m]
0.00 ... 99999.99 [m
3
]
Enter full-scale range value
for the current output (20
mA).
(see explanatory notes)
3.3.5
BAUDRATE
Select 1200 Bd
Baud rate for HART®
communication (do not
change!).
3.3.6 ADDRESS
Enter 0 ... 255 Enter device address.
(for HART® multidrop)
3.3.7 PROTOCOL
Select
HART/KROHNE-PC
Select communications
protocol
3.4 USER
DATA
3.4.1
LANGUAGE
Select GB-USA/D/F/I/E/P/S Select language for the
optional display.
3.4.2
ENTRY CODE 1
Select NO/YES
Switch the access lockout